Skip to main content
NABL Accredited ISO/IEC 17025:2017 • CC-4178 Permanent Lab Calibration + On-Site Field Calibration 26 Scopes

Durometer - Rubber Hardness Calibration Services

Rubber / Shore Hardness Tester (Durometer) - Scales A, B, C, D, DO, E, O, OO, OOO, OOOS, AM, AO, DO, C1L, Depth Indentation & Hardness Testing Machines.

WhatsApp Lab Desk
Calibration Capability (CMC)
Durometer: ± 0.3 Shore | Depth: ± 1.02 µm | Leeb: ± 7.7 HLD | Brinell: ± 1.35% | Rockwell: ± 0.53 HR | Vickers: ± 1.65%.
Primary Master Equipment
Laser Measuring Machine (LMM), Electronic Comparator, Loadcell with Indicator, Standard Hardness Test Blocks.
Governing Standards
ASTM D2240, ASTM A956, IS 1500 : Part 2 : 2021, IS 1586 : Part 2 : 2018, IS 1501 : Part 2 : 2020, ISO 6506-2, ISO 6508-2, ISO 6507-2.
Key Parameters & Applications
Using LMM & Electronic Comparator by Depth Indentation Method as per ASTM D 2240, Loadcell with Indicator by Spring Force Method as per ASTM D2240, and Standard Hardness Test Blocks as per ASTM A956, IS 1500, IS 1586, IS 1501.

Instruments & Parameters Covered (26)

Search or review approved measurement ranges and Calibration & Measurement Capabilities (CMC).

Instrument / Parameter Calibration Range CMC (±) Master / Method
Hardness Tester (Depth Indentation by LMM) 0 to 2 mm 1.02 µm Using LMM by Depth Indentation Method as per ASTM D 2240
Hardness Tester (Depth Indentation by Electronic Comparator) 0 to 2 mm 1.66 µm Using Electronic Comparator by Depth Indentation Method as per ASTM D 2240
Rubber / Shore Hardness Tester (Durometer) - Scales A, B, E, O, OO, OOO, OOOS, AM, AO, DO, C1L 0 Shore to 100 Shore 0.3 Shore Using Loadcell with Indicator by Spring Force Method as per ASTM D2240
Rubber / Shore Hardness Tester (Durometer) - Scales C, D, DO, O, OO, OOO, OOOS, AM, AO, DO, C1L 0 Shore to 100 Shore 0.3 Shore Using Loadcell with Indicator by Spring Force Method as per ASTM D2240
Leeb (Dynamic) Hardness Tester [Permanent Facility] D Scale 7.7 HLD Using Standard Hardness Test Blocks by Indirect Method as per ASTM A956
Brinell Hardness Testing Machine (HBW 10/1000) HBW 10/1000 1.35 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Brinell Hardness Testing Machine (HBW 10/500) HBW 10/500 1.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Brinell Hardness Testing Machine (HBW 10/3000) HBW 10/3000 1.53 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Brinell Hardness Testing Machine (HBW 2.5 / 187.5) HBW 2.5 / 187.5 1.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Brinell Hardness Testing Machine (HBW 2.5 / 62.5) HBW 2.5 / 62.5 1.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Brinell Hardness Testing Machine (HBW 5/750) HBW 5/750 1.88 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1500 : Part 2 : 2021 / ISO 6506-2:2017 / ASTM E10-23, ASTM E110
Leeb (Dynamic) Hardness Tester [Site Facility] D Scale 7.7 HLD Using Standard Hardness Test Blocks by Indirect Method as per ASTM A956
Micro Vickers Hardness Testing Machine (HV 0.3) HV 0.3 2.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 (Part 2) : 2020 / ISO 6507-2 : 2018 / ASTM E384-22 / ASTM E92-23
Micro Vickers Hardness Testing Machine (HV 0.5) HV 0.5 2.77 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 (Part 2) : 2020 / ISO 6507-2 : 2018 / ASTM E384-22 / ASTM E92-23
Micro Vickers Hardness Testing Machine (HV 0.2) HV 0.2 2.72 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 (Part 2) : 2020 / ISO 6507-2 : 2018 / ASTM E384-22 / ASTM E92-23
Micro Vickers Hardness Testing Machine (HV 1) HV 1 2.85 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 (Part 2) : 2020 / ISO 6507-2 : 2018 / ASTM E384-22 / ASTM E92-23
Rockwell Hardness Testing Machine (HR 15N) HR 15N 0.53 HR 15N Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Rockwell Hardness Testing Machine (HR 30N) HR 30N 0.57 HR 30N Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Rockwell Hardness Testing Machine (HR 45N) HR 45N 0.62 HR 45N Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Rockwell Hardness Testing Machine (HRA) HRA 0.54 HRA Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Rockwell Hardness Testing Machine (HRBW) HRBW 0.62 HRBW Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Rockwell Hardness Testing Machine (HRC) HRC 0.53 HRC Using Standard Hardness Test Blocks by Indirect Method as per IS 1586 : Part 2 : 2018 / ISO 6508-2:2023 / ASTM E18-22
Vickers Hardness Testing Machine (HV 10) HV 10 1.68 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 Part 2: 2020 / ISO 6507-2:2018 / ASTM E384-22 / ASTM E92-23
Vickers Hardness Testing Machine (HV 30) HV 30 1.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 Part 2: 2020 / ISO 6507-2:2018 / ASTM E384-22 / ASTM E92-23
Vickers Hardness Testing Machine (HV 50) HV 50 1.65 % Using Standard Hardness Test Blocks by Indirect Method as per IS 1501 Part 2: 2020 / ISO 6507-2:2018 / ASTM E384-22 / ASTM E92-23

Frequently Asked Questions

Are Durometer - Rubber Hardness calibrations NABL accredited?

Yes. Our Durometer - Rubber Hardness laboratory is accredited under ISO/IEC 17025:2017 by NABL (Certificate No: CC-4178), ensuring global acceptance and unbroken traceability to NPL/NIST primary standards.

Do you offer on-site Durometer - Rubber Hardness calibration at client plants?

Yes, Reva Calibration Laboratories deploys mobile metrology teams across Chennai, Tamil Nadu, and PAN-India industrial hubs for on-site Durometer - Rubber Hardness calibration, minimizing equipment downtime.

What is the standard turnaround time for calibration certificates?

Standard turnaround time is 3 to 5 working days for permanent laboratory orders. Express same-day and next-day calibration slots are available upon prior request.

How do I request a quotation for Durometer - Rubber Hardness?

Click the "Request a Quote" button on this page to pre-select Durometer - Rubber Hardness, upload your instrument list (up to 15 MB PDF/Excel), or call +91 98406 72352 for immediate scope verification.

Explore Related Calibration Disciplines & Services

Industrial Turnkey AMC Plans • NABL Lab Setup Consultancy • ISO/IEC 17025:2017 Training Courses • Direct Lab Helpdesk
Call Lab WhatsApp