| Angle Plate - Flatness |
0 to 600 mm |
± 3.8 µm |
Using Plunger Dial, Surface Plate by Comparison Method |
| Box Angle Plate - Flatness |
0 to 600 mm |
± 3.8 µm |
Using Plunger Dial, Surface Plate by Comparison Method |
| Angle Plate - Parallelism |
Up to 600 mm |
± 3.8 µm |
Using Plunger Dial, Surface Plate by Comparison Method |
| Box Angle Plate - Parallelism |
Up to 600 mm |
± 3.8 µm |
Using Plunger Dial, Surface Plate by Comparison Method |
| Angle Plate - Squareness |
Up to 600 mm |
± 3.8 µm |
Using Cylindrical Square, Surface Plate, Gauge Block by Comparison Method |
| Box Angle Plate - Squareness |
Up to 600 mm |
± 3.8 µm |
Using Cylindrical Square, Surface Plate, Gauge Block by Comparison Method |
| Bridge Cam Gauge - Angle |
0° to 60° |
± 2.47 minute of arc |
Using Profile Projector by Direct Method |
| Bridge Cam Gauge - Linear |
Up to 60 mm |
± 4.6 µm |
Using Profile Projector by Direct Method |
| Brinell Ball Indenter - Linear |
Up to 10 mm |
± 2.61 µm |
Using Profile Projector by Direct Method |
| Caliper - Vernier / Dial / Digital / Groove (L.C.: 0.01 mm & Coarser) |
0 to 2000 mm |
± 0 to 1000 mm: 10.55 µm | 0 to 2000 mm: 15.8 µm |
Using Long Slip Gauge, Gauge Block by Comparison Method |
| Caliper - Vernier / Dial / Digital / Groove (L.C.: 0.01 mm & Coarser) |
0 to 600 mm |
± 0 to 300 mm: 6.99 µm | 0 to 600 mm: 8.8 µm |
Using Step Gauge (Caliper Checker), Gauge Block by Comparison Method |
| Chamfer Gauge Set - Angle |
0° to 90° |
± 10 minute of arc |
Using Profile Projector by Comparison Method |
| Chamfer Gauge Set - Length |
0 to 50 mm |
± 11.4 µm |
Using Profile Projector by Comparison Method |
| Clinometer (L.C.: 0.5' & Coarser) |
0° to 360° |
± 1.49 minute of arc |
Using Sine Bar, Gauge Block Set by Comparison Method as per MOY/SIRA/36 |
| Coating Thickness Gauge (L.C.: 0.001 mm) |
0.01 mm to 2 mm |
± 0.6 µm |
Using Foils by Comparison Method |
| DFT Gauge (L.C.: 0.001 mm) |
0.01 mm to 2 mm |
± 0.6 µm |
Using Foils by Comparison Method |
| DFT Meter (L.C.: 0.001 mm) |
0.01 mm to 2 mm |
± 0.6 µm |
Using Foils by Comparison Method |
| Coating Thickness Gauge (L.C.: 0.01 mm or Coarser) |
> 2 mm to 10 mm |
± 5.1 µm |
Using Foils by Comparison Method |
| DFT Gauge (L.C.: 0.01 mm or Coarser) |
> 2 mm to 10 mm |
± 5.1 µm |
Using Foils by Comparison Method |
| DFT Meter (L.C.: 0.01 mm or Coarser) |
> 2 mm to 10 mm |
± 5.1 µm |
Using Foils by Comparison Method |
| Combination Set (L.C.: 1°) |
0° - 90° - 0° |
± 19.29 minute of arc |
Using Profile Projector by Direct Method |
| Comparator Stand - Flatness |
Up to 50 mm |
± 0.3 µm |
Using LVDT Electronic Comparator and Optical Flat by Direct Method |
| Cone Master - Angle |
55° & 60° |
± 3 minute of arc |
Using Profile Projector by Direct Method |
| Crimping Tool - Angular |
0° to 180° |
± 6 minute of arc |
Using Profile Projector by Comparison Method |
| Crimping Tool - Linear |
Up to 50 mm |
± 7 µm |
Using Profile Projector by Comparison Method |
| Cross Hatch Cutter - Angle |
45° |
± 3.6 minute of arc |
Using Profile Projector by Direct Method |
| Cross Hatch Cutter - Pitch |
Up to 4 mm |
± 2.8 µm |
Using Profile Projector by Direct Method |
| Cube Mould - Linear |
Up to 1000 mm |
± 26.5 µm |
Using Digital Caliper by Comparison Method as per IS 10086 |
| Rectangular Mould - Linear |
Up to 1000 mm |
± 26.5 µm |
Using Digital Caliper by Comparison Method as per IS 10086 |
| Cylindrical Mould - Linear |
Up to 1000 mm |
± 26.5 µm |
Using Digital Caliper by Comparison Method as per IS 10086 |
| Degree Protractor (L.C.: 0.1° & Coarser) |
0° to 180° |
± 5.16 minute of arc |
Using Profile Projector by Direct Method |
| Depth Gauge - Vernier / Dial / Hook / Digital (L.C.: 0.01 mm & Coarser) |
0 to 1000 mm |
± 0 to 300 mm: 4.3 µm | > 300 mm to 1000 mm: 8.6 µm |
Using Slip Gauge and Long Slip Gauge by Comparison Method |
| Depth Gauge - Vernier / Dial / Hook / Digital (L.C.: 0.01 mm & Coarser) |
0 to 300 mm |
± 7.39 µm |
Using Step Gauge (Depth Micro - Checker) & Gauge Block by Comparison Method |
| Depth Micrometer - Dial / Digital / Analog (L.C.: 0.001 mm & Coarser) |
0 to 300 mm |
± 0 to 100 mm: 1.05 µm | > 100 mm to 300 mm: 2.04 µm |
Using Slip Gauge and Long Slip Gauge by Comparison Method as per JIS B 7544 |
| Depth Micrometer - Dial / Digital / Analog (L.C.: 0.001 mm & Coarser) |
0 to 300 mm |
± 3.86 µm |
Using Step Gauge (Depth Micro - Checker), Slip Gauge by Comparison Method as per JIS B 7544 |
| Dial Bore Gauge - Transmission Error (L.C.: 0.001 mm & Coarser) [Stem / Split Type] |
Up to 1.5 mm |
± 2.01 µm |
Using Electronic Dial Calibration Tester & Dial Indicator by Comparison Method as per JIS B 7515 |
| Bore Gauge - Transmission Error (L.C.: 0.001 mm & Coarser) [Stem / Split Type] |
Up to 1.5 mm |
± 2.01 µm |
Using Electronic Dial Calibration Tester & Dial Indicator by Comparison Method as per JIS B 7515 |
| Dial Bore Gauge - Transmission Error (L.C.: 0.001 mm & Coarser) [Stem / Split Type] |
Up to 2 mm |
± 0.78 µm |
Using Length Measuring Machine by Direct Method as per JIS B 7515 |
| Bore Gauge - Transmission Error (L.C.: 0.001 mm & Coarser) [Stem / Split Type] |
Up to 2 mm |
± 0.78 µm |
Using Length Measuring Machine by Direct Method as per JIS B 7515 |
| Dial Caliper Gauge - Dial / Digital [Internal Type] (L.C.: 0.005 mm & Coarser) |
0 to 100 mm |
± 0 mm to 50 mm: 2.82 µm | > 50 mm to 100 mm: 2.85 µm |
Using Gauge Block, Slip Gauge Accessories Set by Comparison Method |
| Groove Dial Gauge - Dial / Digital [Internal Type] (L.C.: 0.005 mm & Coarser) |
0 to 100 mm |
± 0 mm to 50 mm: 2.82 µm | > 50 mm to 100 mm: 2.85 µm |
Using Gauge Block, Slip Gauge Accessories Set by Comparison Method |
| Dial Caliper Gauge - Dial / Digital [Internal Type] (L.C.: 0.005 mm & Coarser) |
> 100 mm to 200 mm |
± 2.92 µm |
Using Long Gauge Blocks, Slip Gauge Accessories Set by Comparison Method |
| Groove Dial Gauge - Dial / Digital [Internal Type] (L.C.: 0.005 mm & Coarser) |
> 100 mm to 200 mm |
± 2.92 µm |
Using Long Gauge Blocks, Slip Gauge Accessories Set by Comparison Method |
| Dial Comparator (L.C.: 0.0001 mm) |
0 to 50 mm |
± 0.66 µm |
Using Length Measuring Machine by Comparison Method |
| Plunger Type Dial Gauge (L.C.: 0.0001 mm) |
0 to 50 mm |
± 0.66 µm |
Using Length Measuring Machine by Comparison Method |
| Digital Indicator (L.C.: 0.0001 mm) |
0 to 50 mm |
± 0.66 µm |
Using Length Measuring Machine by Comparison Method |
| Dial Snap Gauge - Digital / Dial (L.C.: 0.001 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 0.98 µm | > 50 mm to 100 mm: 1.07 µm |
Using Gauge Blocks by Comparison Method |
| Dial Snap Micrometer - Digital / Dial (L.C.: 0.001 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 0.98 µm | > 50 mm to 100 mm: 1.07 µm |
Using Gauge Blocks by Comparison Method |
| Elongation Gauge - Length |
14.7 mm to 81 mm |
± 2.97 µm |
Using Profile Projector by Direct Method |
| Engineer Parallel - Parallelism |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block, Dial Indicator / Precision Level by Comparison Method |
| Parallel Block - Parallelism |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block, Dial Indicator / Precision Level by Comparison Method |
| Engineer Parallel - Squareness |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block & Dial Indicator, Cylindrical Square by Comparison Method |
| Parallel Block - Squareness |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block & Dial Indicator, Cylindrical Square by Comparison Method |
| Engineer Parallel - Width & Height Variation of Matched Pairs |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block & LVDT by Comparison Method |
| Parallel Block - Width & Height Variation of Matched Pairs |
Up to 300 mm |
± 2.18 µm |
Using Surface Plate, Gauge Block & LVDT by Comparison Method |
| Engineer Square - Flatness |
Up to 1000 mm |
± 4.4 µm |
Using Surface Plate & Dial Gauge by Comparison Method |
| Try Square - Flatness |
Up to 1000 mm |
± 4.4 µm |
Using Surface Plate & Dial Gauge by Comparison Method |
| Engineer Square - Parallelism |
Up to 1000 mm |
± 4.4 µm |
Using Surface Plate & Dial Gauge by Comparison Method |
| Try Square - Parallelism |
Up to 1000 mm |
± 4.4 µm |
Using Surface Plate & Dial Gauge by Comparison Method |
| Engineer Square - Squareness |
Up to 1000 mm |
± 4.4 µm |
Using Granite Square, Surface Plate & Dial Gauge by Comparison Method |
| Try Square - Squareness |
Up to 1000 mm |
± 4.4 µm |
Using Granite Square, Surface Plate & Dial Gauge by Comparison Method |
| Extensometer (L.C.: 0.001 mm & Coarser) |
0 to 5 mm |
± 2.95 µm |
Using Extensometer Calibration Stand with DRO & Digital Caliper by Comparison Method as per ASTM E83 & ISO 9513 |
| External Micrometer - Flange / Ball / Blade / Point / Disc / Groove / Spline / Sheet / Pitch (Analog / Digital) (L.C.: 0.0001 mm & Coarser) |
0 to 100 mm |
± 0 to 25 mm: 0.54 µm | 0 to 100 mm: 0.78 µm |
Using Gauge Block Set by Comparison Method |
| External Micrometer - Flange / Ball / Blade / Point / Disc / Groove / Spline / Sheet / Pitch Type (Analog / Digital) (L.C.: 0.001 mm & Coarser) |
100 to 600 mm |
± > 100 mm to 300 mm: 2.51 µm | > 300 mm to 600 mm: 5.53 µm |
Using Gauge Blocks, Length Bars and Long Slip Gauge Set by Comparison Method |
| External Micrometer - Flange / Ball / Blade / Point / Disc / Groove / Spline / Sheet / Pitch Type (Analog / Digital) (L.C.: 0.001 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 0.54 µm | > 50 mm to 100 mm: 0.75 µm |
Using Gauge Blocks by Comparison Method |
| External Micrometer - Flange / Ball / Blade / Point / Disc / Groove / Spline / Sheet / Pitch Type (Analog / Digital) (L.C.: 0.01 mm & Coarser) |
> 1200 mm to 1800 mm |
± 12.46 µm |
Using Gauge Blocks, Length Bars and Long Slip Gauge Set by Comparison Method |
| External Micrometer - Flange / Ball / Blade / Point / Disc / Groove / Spline / Sheet / Pitch Type (Analog / Digital) (L.C.: 0.01 mm & Coarser) |
600 mm to 1200 mm |
± 11.57 µm |
Using Gauge Blocks, Length Bars by Comparison Method |
| Feeler Gauge Set |
0.01 mm to 2 mm |
± 0.6 µm |
Using Electronic Comparator & Comparator Stand by Comparison Method |
| Thickness Gauge |
0.01 mm to 2 mm |
± 0.6 µm |
Using Electronic Comparator & Comparator Stand by Comparison Method |
| Thickness Standard |
0.01 mm to 2 mm |
± 0.6 µm |
Using Electronic Comparator & Comparator Stand by Comparison Method |
| Feeler Gauge Set |
0.01 mm to 2 mm |
± 0.6 µm |
Using Length Measuring Machine by Direct Method |
| Thickness Gauge |
0.01 mm to 2 mm |
± 0.6 µm |
Using Length Measuring Machine by Direct Method |
| Thickness Standard |
0.01 mm to 2 mm |
± 0.6 µm |
Using Length Measuring Machine by Direct Method |
| Flush Pin |
0.2 mm to 200 mm |
± 3.2 µm |
Using Slip Gauge, Electronic Comparator, Comparator Stand by Comparison Method |
| Flush Pin Gauge (Step Limit) |
Up to 5 mm |
± 1.78 µm |
Using Electronic Comparator & Gauge Block by Comparison Method |
| Gear Tooth Vernier (L.C.: 0.01 mm) |
Up to 50 mm |
± 6.7 µm |
Using Gauge Blocks by Comparison Method |
| Groove Micrometer (L.C.: 0.01 mm & Coarser) |
0 to 100 mm |
± 3.47 µm |
Using Gauge Block by Comparison Method |
| Hegman Gauge |
0 to 100 µm |
± 0.94 µm |
Using Electronic Comparator with DRO by Direct Method |
| Height Gauge - Vernier / Dial / Electronic / Digital (L.C.: 0.01 mm & Coarser) |
0 to 600 mm |
± 0 to 300 mm: 3.18 µm | 0 to 600 mm: 7.4 µm |
Using Caliper Checker by Comparison Method |
| Height Gauge - Vernier / Dial / Electronic / Digital (L.C.: 0.01 mm & Coarser) |
0 to 1000 mm |
± 0 to 600 mm: 7.4 µm | > 600 mm to 1000 mm: 7.8 µm |
Using Gauge Block, Long Slip Gauge by Comparison Method |
| Hi - Lo Gauge - Linear |
Up to 100 mm |
± 2.95 µm |
Using Profile Projector & Gauge Blocks by Direct Method |
| Impact Ball - Diameter |
Up to 50 mm |
± 0.14 µm |
Using Length Measuring Machine by Direct Method |
| Steel Sphere - Diameter |
Up to 50 mm |
± 0.14 µm |
Using Length Measuring Machine by Direct Method |
| Sphere - Diameter |
Up to 50 mm |
± 0.14 µm |
Using Length Measuring Machine by Direct Method |
| Reference Sphere - Diameter |
Up to 50 mm |
± 0.14 µm |
Using Length Measuring Machine by Direct Method |
| Inclinometer - Analog / Digital (L.C.: 0.01° & Coarser) |
0° to 180° |
± 6.1 minute of arc |
Using Sine Bar & Gauge Block Set by Comparison Method |
| Inside Micrometer (L.C.: 0.001 mm & Coarser) |
5 mm to 25 mm |
± 2 µm |
Using Gauge Blocks & Gauge Block Accessories Set by Comparison Method |
| Internal Micrometer (Caliper Type) (L.C.: 0.001 mm & Coarser) |
5 mm to 25 mm |
± 2 µm |
Using Gauge Blocks & Gauge Block Accessories Set by Comparison Method |
| Inside Micrometer (L.C.: 0.01 mm & Coarser) |
13 mm to 1000 mm |
± 5.5 µm |
Using Gauge Blocks, Long Slip Gauge Set, Electronic Comparator with Stand / Lever Dial Indicator by Comparison Method |
| Internal Micrometer with Connecting Rod (Stick / Jaw / Tubular Type) (L.C.: 0.01 mm & Coarser) |
13 mm to 1000 mm |
± 5.5 µm |
Using Gauge Blocks, Long Slip Gauge Set, Electronic Comparator with Stand / Lever Dial Indicator by Comparison Method |
| Joint Finger Probe - Length |
0 to 300 mm |
± 20.5 µm |
Using Digital Caliper by Comparison Method |
| Keyway Symmetry Gauge - ID |
0 to 100 mm |
± 10.41 µm |
Using Profile Projector by Comparison Method |
| Keyway Symmetry Gauge - OD |
0 to 100 mm |
± 10.41 µm |
Using Profile Projector by Comparison Method |
| Length Gauge - Linear |
0 to 200 mm |
± 2.48 µm |
Using Gauge Blocks by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.001 mm & Coarser) |
0 to 0.14 mm |
± 2 µm |
Using Electronic Dial Calibration Tester by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.001 mm & Coarser) |
0 to 0.14 mm |
± 0.68 µm |
Using Length Measuring Machine by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.002 mm & Coarser) |
0 to 0.6 mm |
± 2.91 µm |
Using Electronic Dial Calibration Tester by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.002 mm & Coarser) |
0 to 0.6 mm |
± 0.92 µm |
Using Length Measuring Machine by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.01 mm & Coarser) |
0 to 2 mm |
± 0.96 µm |
Using Length Measuring Machine by Comparison Method |
| Lever Type Dial Gauge (L.C.: 0.01 mm & Coarser) |
0 to 2 mm |
± 2 µm |
Using Electronic Dial Calibration Tester by Comparison Method |
| Limit Gauge - Geometrical Dimension |
Up to 300 mm |
± 4.07 µm |
Using Vernier Caliper & Profile Projector by Direct Method |
| Profile Gauge - Geometrical Dimension |
Up to 300 mm |
± 4.07 µm |
Using Vernier Caliper & Profile Projector by Direct Method |
| Special Gauge - Geometrical Dimension |
Up to 300 mm |
± 4.07 µm |
Using Vernier Caliper & Profile Projector by Direct Method |
| Industrial Inspection Gauge (Height, Length, Width, Diameter, Depth, Gap, Knife Edge, Inspection Jig & Fixture) - Geometrical Dimension |
Up to 300 mm |
± 4.07 µm |
Using Vernier Caliper & Profile Projector by Direct Method |
| Limit Gauge - Angle |
Up to 360° |
± 3.2 minute of arc |
Using Profile Projector by Direct Method |
| Profile Gauge - Angle |
Up to 360° |
± 3.2 minute of arc |
Using Profile Projector by Direct Method |
| Special Gauge - Angle |
Up to 360° |
± 3.2 minute of arc |
Using Profile Projector by Direct Method |
| Industrial Inspection Gauge - Angle |
Up to 360° |
± 3.2 minute of arc |
Using Profile Projector by Direct Method |
| Master Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.77 µm | > 2 mm to 10 mm: 0.93 µm |
Using Length Measuring Machine by Comparison Method |
| Thickness Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.77 µm | > 2 mm to 10 mm: 0.93 µm |
Using Length Measuring Machine by Comparison Method |
| Standard Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.77 µm | > 2 mm to 10 mm: 0.93 µm |
Using Length Measuring Machine by Comparison Method |
| Master Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.45 µm | > 2 mm to 10 mm: 0.8 µm |
Using Electronic Comparator with DRO by Comparison Method |
| Thickness Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.45 µm | > 2 mm to 10 mm: 0.8 µm |
Using Electronic Comparator with DRO by Comparison Method |
| Standard Foil |
0.005 to 10 mm |
± 0.005 mm to 2 mm: 0.45 µm | > 2 mm to 10 mm: 0.8 µm |
Using Electronic Comparator with DRO by Comparison Method |
| Measuring Pin |
0.1 mm to 45 mm |
± 0.59 µm |
Using LMM by Comparison Method |
| Pin Gauge |
0.1 mm to 45 mm |
± 0.59 µm |
Using LMM by Comparison Method |
| Cylindrical Measuring Pin |
0.1 mm to 45 mm |
± 0.59 µm |
Using LMM by Comparison Method |
| Measuring Steel Rule (L.C.: 0.5 mm & Coarser) |
Up to 2000 mm |
± 115.8 x sqrt (L) µm, where L is in mm |
Using Tape & Scale Measuring Machine by Comparison Method |
| Steel Scale (L.C.: 0.5 mm & Coarser) |
Up to 2000 mm |
± 115.8 x sqrt (L) µm, where L is in mm |
Using Tape & Scale Measuring Machine by Comparison Method |
| Plastic Rule (L.C.: 0.5 mm & Coarser) |
Up to 2000 mm |
± 115.8 x sqrt (L) µm, where L is in mm |
Using Tape & Scale Measuring Machine by Comparison Method |
| Measuring Tape (L.C.: 0.5 mm & 1 mm & 0.01 mm & Coarser) |
Up to 200 m |
± 110.4 x sqrt (L) µm, where L is in mm |
Using Tape & Scale Measuring Machine by Comparison Method |
| PIE Tape (L.C.: 0.5 mm & 1 mm & 0.01 mm & Coarser) |
Up to 200 m |
± 110.4 x sqrt (L) µm, where L is in mm |
Using Tape & Scale Measuring Machine by Comparison Method |
| Micro Dial Comparator (L.C.: 0.1 µm & Coarser) |
(±) 0.05 mm |
± 0.55 µm |
Using Electronic Probe with DRO by Comparison Method |
| Micrometer Head - Digital (L.C.: 0.0002 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 0.76 µm | > 50 mm to 100 mm: 0.86 µm |
Using LMM by Comparison Method |
| Micrometer Head - Digital (L.C.: 0.0002 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 1.76 µm | > 50 mm to 100 mm: 1.91 µm |
Using Electronic Probe with DRO by Comparison Method |
| Micrometer Head - Digital (L.C.: 0.0002 mm & Coarser) |
0 to 25 mm |
± 1.74 µm |
Using Electronic Comparator by Comparison Method |
| Micrometer Setting Standard |
> 100 mm to 300 mm |
± 2.1 µm |
Using Length Measuring Machine, Long Gauge Block by Comparison Method |
| Micrometer Setting Standard |
25 mm to 100 mm |
± 0.9 µm |
Using Slip Gauge & Electronic Probe by Comparison Method |
| Micrometer Setting Standard |
300 mm to 1000 mm |
± 6.8 µm |
Using Long Slip Gauge & LVDT Probe by Comparison Method |
| Micrometer Setting Standard |
50 mm to 100 mm |
± 2.1 µm |
Using Length Measuring Machine, Slip Gauge & Long Gauge Block by Comparison Method |
| Notch Profile Template - Angular |
0° to 45° |
± 2.4 minute of arc |
Using Profile Projector by Direct Method |
| Notch Profile Template - Linear |
Up to 75 mm |
± 2.61 µm |
Using Profile Projector by Direct Method |
| Paddle Gauge [Plain Plug Gauge] |
1 mm to 300 mm |
± 2.1 µm |
Using Electronic Comparator & Comparator Stand by Direct Method |
| Paddle Gauge [Plain Plug Gauge] |
1 mm to 50 mm |
± 1.1 µm |
Using Electronic Comparator with DRO by Comparison Method |
| PCD Gauge - Linear |
2.5 mm to 100 mm |
± 5.1 µm |
Using 2D Electronic Height Gauge by Comparison Method |
| Penetrometer (Plunger Dial Gauge) (L.C.: 0.1 mm & Coarser) |
0 to 40 mm |
± 61.1 µm |
Using Gauge Block Set by Direct Method |
| Pistol Caliper (L.C.: 0.001 mm & Coarser) |
0 to 100 mm |
± 1.91 µm |
Using Gauge Block, Long Gauge Block by Comparison Method |
| Pistol Caliper (L.C.: 0.1 mm & Coarser) |
0 to 250 mm |
± 61.1 µm |
Using Gauge Block, Long Gauge Block by Comparison Method |
| OD Caliper (L.C.: 0.1 mm & Coarser) |
0 to 250 mm |
± 61.1 µm |
Using Gauge Block, Long Gauge Block by Comparison Method |
| OD Groove Dial Gauge (L.C.: 0.1 mm & Coarser) |
0 to 250 mm |
± 61.1 µm |
Using Gauge Block, Long Gauge Block by Comparison Method |
| Pitch Micrometer - Angular (L.C.: 0.001 mm & Coarser) |
45° |
± 2.47 minute of arc |
Using Profile Projector by Direct Method |
| Pitch Micrometer - Linear (L.C.: 0.001 mm & Coarser) |
0 to 300 mm |
± 0 to 100 mm: 0.86 µm | > 100 mm to 300 mm: 1.16 µm |
Using Gauge Blocks by Comparison Method |
| Plain Gap Gauge - Fixed / Adjustable |
> 100 mm to 200 mm |
± 1.95 µm |
Using Long Gauge Blocks by Comparison Method |
| Hub Gauge - Fixed / Adjustable |
> 100 mm to 200 mm |
± 1.95 µm |
Using Long Gauge Blocks by Comparison Method |
| Snap Gauge - Fixed / Adjustable |
> 100 mm to 200 mm |
± 1.95 µm |
Using Long Gauge Blocks by Comparison Method |
| Plain Gap Gauge - Fixed / Adjustable |
> 200 mm to 300 mm |
± 2.2 µm |
Using Long Gauge Blocks by Comparison Method |
| Hub Gauge - Fixed / Adjustable |
> 200 mm to 300 mm |
± 2.2 µm |
Using Long Gauge Blocks by Comparison Method |
| Snap Gauge - Fixed / Adjustable |
> 200 mm to 300 mm |
± 2.2 µm |
Using Long Gauge Blocks by Comparison Method |
| Plain Gap Gauge - Fixed / Adjustable |
1 to 100 mm |
± 1 mm to 50 mm: 0.73 µm | > 50 mm to 100 mm: 1.02 µm |
Using Gauge Blocks by Comparison Method |
| Hub Gauge - Fixed / Adjustable |
1 to 100 mm |
± 1 mm to 50 mm: 0.73 µm | > 50 mm to 100 mm: 1.02 µm |
Using Gauge Blocks by Comparison Method |
| Snap Gauge - Fixed / Adjustable |
1 to 100 mm |
± 1 mm to 50 mm: 0.73 µm | > 50 mm to 100 mm: 1.02 µm |
Using Gauge Blocks by Comparison Method |
| Plain Plug Gauge |
1 to 300 mm |
± 1 mm to 50 mm: 0.9 µm | > 50 mm to 100 mm: 1.5 µm | > 100 mm to 300 mm: 2.5 µm |
Using LMM by Comparison Method |
| Air Plug Gauge |
1 to 300 mm |
± 1 mm to 50 mm: 0.9 µm | > 50 mm to 100 mm: 1.5 µm | > 100 mm to 300 mm: 2.5 µm |
Using LMM by Comparison Method |
| Plain Ring Gauge |
3 to 300 mm |
± 3 mm to 50 mm: 1.31 µm | > 50 mm to 100 mm: 1.7 µm | > 100 mm to 150 mm: 1.8 µm | > 150 mm to 215 µm: 2.1 µm | > 215 mm to 300 mm: 2.2 µm |
Using LMM by Comparison Method |
| Setting Ring Gauge |
3 to 300 mm |
± 3 mm to 50 mm: 1.31 µm | > 50 mm to 100 mm: 1.7 µm | > 100 mm to 150 mm: 1.8 µm | > 150 mm to 215 µm: 2.1 µm | > 215 mm to 300 mm: 2.2 µm |
Using LMM by Comparison Method |
| Air Ring Gauge |
3 to 300 mm |
± 3 mm to 50 mm: 1.31 µm | > 50 mm to 100 mm: 1.7 µm | > 100 mm to 150 mm: 1.8 µm | > 150 mm to 215 µm: 2.1 µm | > 215 mm to 300 mm: 2.2 µm |
Using LMM by Comparison Method |
| Master Ring Gauge |
50 to 300 mm |
± > 50 mm to 100 mm: 1.7 µm | > 100 mm to 150 mm: 1.8 µm | > 150 mm to 215 µm: 2.1 µm | > 215 mm to 300 mm: 2.2 µm |
Using LMM by Comparison Method |
| Plunger Type Dial Gauge - Analog / Digital (L.C.: 0.001 mm & Coarser) |
0 to 50 mm |
± 0.35 µm |
Using LMM by Comparison Method |
| Dial Indicator - Analog / Digital (L.C.: 0.001 mm & Coarser) |
0 to 50 mm |
± 0.35 µm |
Using LMM by Comparison Method |
| Plunger Type Dial Gauge - Analog / Digital (L.C.: 0.01 mm & Coarser) |
0 to 100 mm |
± 3.12 µm |
Using Length Measuring Machine by Comparison Method |
| Dial Indicator - Analog / Digital (L.C.: 0.01 mm & Coarser) |
0 to 100 mm |
± 3.12 µm |
Using Length Measuring Machine by Comparison Method |
| Plunger Type Dial Gauge - Analog / Digital (L.C.: 0.001 mm & Coarser) |
0 to 25 mm |
± 0 to 5 mm: 2.5 µm | 0 to 25 mm: 2.51 µm |
Using Electronic Dial Calibration Tester & Gauge Blocks by Comparison Method |
| Dial Indicator - Analog / Digital (L.C.: 0.001 mm & Coarser) |
0 to 25 mm |
± 0 to 5 mm: 2.5 µm | 0 to 25 mm: 2.51 µm |
Using Electronic Dial Calibration Tester & Gauge Blocks by Comparison Method |
| Proximity Sensor (L.C.: 1 mm & Coarser) |
0 to 1000 mm |
± 446.1 µm |
Using Length Bar & Gauge Block Set by Direct Method |
| Laser Distance Meter (L.C.: 1 mm & Coarser) |
0 to 1000 mm |
± 446.1 µm |
Using Length Bar & Gauge Block Set by Direct Method |
| Radius Gauge - Concave / Convex |
0.6 mm to 40 mm |
± 4.9 µm |
Using Profile Projector by Direct Method |
| Radius Template - Concave / Convex |
0.6 mm to 40 mm |
± 4.9 µm |
Using Profile Projector by Direct Method |
| Receiver Gauge - Linear |
Up to 200 mm |
± 4.7 µm |
Using 2D Electronic Height Gauge by Direct Method |
| Riser Block - Linear |
Up to 300 mm |
± 2.44 µm |
Using Lever Dial Probe with Indicator, Check Master by Comparison Method |
| Rockwell Diamond Cone Indenter - Angular |
Up to 120° |
± 2.5 minute of arc |
Using Profile Projector by Direct Method as per ASTM E18 |
| Rockwell Diamond Cone Indenter - Radius |
Up to 0.2 mm |
± 3.89 µm |
Using Profile Projector by Direct Method as per ASTM E18 |
| Spirit Level - Base Length (Up to 300 mm) |
(±) 2 mm |
± 6.07 µm/m |
Using Electronic Level And Tilting Table by Comparison Method |
| Frame Level - Base Length (Up to 300 mm) |
(±) 2 mm |
± 6.07 µm/m |
Using Electronic Level And Tilting Table by Comparison Method |
| Coincidence Level (Type 1 & 2 & 3) (L.C.: 10 µm / m & Coarser) - Base Length (Up to 300 mm) |
(±) 2 mm |
± 6.07 µm/m |
Using Electronic Level And Tilting Table by Comparison Method |
| Spline Plug Gauge - Dimension between Measuring Pins |
10 to 300 mm |
± 10 mm to 100 mm: 1.02 µm | > 100 mm to 300 mm: 1.9 µm |
Using LMM & Measuring Pins by Comparison Method as per DIN 5480-1 |
| Spline Plug Gauge - Dimension between Measuring Pins |
10 mm to 100 mm |
± 2.58 µm |
Using FCDM & Measuring Pins by Comparison Method as per DIN 5480-1 |
| Spline Ring Gauge - Dimension between Measuring Pins |
10 mm to 100 mm |
± 2.2 µm |
Using LMM & Measuring Pins by Comparison Method as per DIN 5480-1 |
| Step Thickness Gauge Block |
1 mm to 300 mm |
± 21.36 µm |
Using LVDT and Long Gauge Block by Comparison Method |
| Step Thickness Gauge Block |
1 mm to 20 mm |
± 1.11 µm |
Using Electronic Comparator with DRO and Gauge Block Set by Comparison Method |
| Straight Edge - Parallelism |
Up to 3000 mm |
± 5.4 µm |
Using Surface Plate, LVDT & Dial Gauge with Stand by Comparison Method |
| Straight Edge - Straightness |
Up to 3000 mm |
± 5.4 µm |
Using Surface Plate & Electronic Level by Comparison Method |
| Surface Plate - Granite / Cast Iron |
0 to 3000 mm |
± Up to 3000 x 3000 mm: 0.53 x sqrt ((L + W) / 100) µm |
Using Electronic Level by Comparison Method |
| Taper Plain Plug Gauge - Diameter |
> 100 mm to 300 mm |
± 2.09 µm |
Using LMM, Gauge Blocks & Measuring Pin by Comparison Method |
| Taper Plain Plug Gauge - Diameter |
6 mm to 100 mm |
± 1.34 µm |
Using LMM, Gauge Blocks & Measuring Pins by Comparison Method |
| Taper Plain Plug Gauge - Taper Angle |
Up to 18° |
± 25 second of arc |
Using LMM, Gauge Block & Measuring Pin by Comparison Method |
| Taper Plain Plug Gauge - Taper Angle |
Up to 28° |
± 31 second of arc |
Using LMM, Gauge Block & Measuring Pin by Comparison Method |
| Taper Plain Ring Gauge - Diameter |
6 to 300 mm |
± 6 mm to 100 mm: 1.5 µm | > 100 mm to 300 mm: 1.74 µm |
Using LMM by Comparison Method |
| Taper Plain Ring Gauge - Taper Angle |
Up to 18° |
± 25 second of arc |
Using LMM by Comparison Method |
| Taper Plain Ring Gauge - Taper Angle |
Up to 28° |
± 31 second of arc |
Using LMM by Comparison Method |
| Taper Scale (L.C.: 0.1 mm & Coarser) |
0.1 mm to 60 mm |
± 2.96 µm |
Using Profile Projector by Comparison Method |
| Taper Bore Gauge (L.C.: 0.1 mm & Coarser) |
0.1 mm to 60 mm |
± 2.96 µm |
Using Profile Projector by Comparison Method |
| Taper Thread Plug Gauge - Effective Diameter |
3 to 100 mm |
± 3 mm to 50 mm: 3.4 µm | > 50 mm to 100 mm: 4 µm |
Using FCDM & Thread Measuring Wire by Comparison Method |
| Taper Thread Plug Gauge - Effective Diameter |
3 to 300 mm |
± 3 mm to 50 mm: 1.01 µm | > 50 mm to 100 mm: 1 µm | > 100 mm to 300 mm: 2 µm |
Using LMM & Thread Measuring Wire by Comparison Method |
| Test Sieve - Perforated Plate / Wire Cloth |
0.010 mm to 5 mm |
± 4 µm |
Using Profile Projector by Direct Method |
| Thickness Gauge (Flakiness Gauge) |
4.89 mm to 33.9 mm |
± 2.97 µm |
Using Profile Projector by Direct Method |
| Thickness Gauge - Digital / Dial (L.C.: 0.001 mm & Coarser) |
0 to 100 mm |
± 0 to 50 mm: 0.98 µm | 0 to 100 mm: 1.07 µm |
Using Gauge Blocks by Comparison Method |
| Thread Measuring Prism - Angle |
Up to 45° |
± 3.46 minute of arc |
Using Profile Projector by Direct Method |
| Thread Measuring Prism - Linear |
3 mm to 6 mm |
± 1.1 µm |
Using Electronic Comparator & Comparator Stand by Comparison Method |
| Thread Measuring Prism - Linear |
3 mm to 6 mm |
± 1.17 µm |
Using Profile Projector by Comparison Method |
| Thread Measuring Wire |
0.17 mm to 6.35 mm |
± 0.25 µm |
Using LMM by Direct Method |