| Instrument / Parameter | Calibration Range | CMC (±) | Master / Method |
|---|---|---|---|
| Electronic Balance - Accuracy Class I and Coarser (Readability: 0.0001 mg) [Permanent] | 0 to 2.1 g / 0 to 6.1 g | 0 to 2.1 g: ± 0.002 mg | 0 to 6.1 g: ± 0.004 mg | Using E1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class I and Coarser (Readability: 0.001 mg) [Permanent] | 0 to 22 g | ± 0.005 mg | Using E1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class I and Coarser (Readability: 0.01 mg) [Permanent] | 0 to 50 g / 0 to 220 g / 0 to 1000 g | 0 to 50 g: ± 0.01 mg | 0 to 220 g: ± 0.03 mg | 0 to 1000 g: ± 0.15 mg | Using E1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 1 mg) [Permanent - 20 kg] | 0 to 20 kg | ± 3.5 mg | Using E1, E2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 1 mg) [Permanent - 5 kg] | 0 to 5 kg | ± 1 mg | Using E1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 10 mg) [Permanent] | 0 to 6.2 kg | ± 6 mg | Using E1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 100 mg) [Permanent - 36 kg] | 0 to 36 kg | ± 62 mg | Using E1, E2, F1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 100 mg) [Permanent - 75 kg] | 0 to 75 kg | ± 300 mg | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class II and Coarser (Readability: 5 mg) [Permanent] | 0 to 40 kg | ± 10 mg | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class III and Coarser (Readability: 0.1 g) [Permanent] | 0 to 50 kg | ± 100 mg | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class III and Coarser (Readability: 1 g) [Permanent] | 0 to 60 kg | ± 0.6 g | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class III and Coarser (Readability: 2 g) [Permanent] | 0 to 100 kg / 0 to 60 kg | ± 1.2 g | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Weighing Scale & Balance (Accuracy Class IV, Readability: 50 g) [Permanent] | 0 to 1000 kg | ± 62 g | Using F2, M1 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class IIII (Readability: 100 g) [Permanent] | 0 to 2000 kg | ± 62 g | Using F2, M1 Class Weights by Comparison Method as per OIML R 76-1 |
| Weighing Scale & Balance (Accuracy Class III, Readability: 1 g) [Permanent] | 0 to 150 kg | ± 2 g | Using E2, F1 & F2 Class Standard Weights as per OIML R 76-1 |
| Electronic Balance - Accuracy Class IIII (Readability: 10 g) [Permanent - 300 kg] | 0 to 300 kg | ± 13 g | Using E2, F1, F2 Class Weights by Comparison Method as per OIML R 76-1 |
| Electronic Balance - Accuracy Class IIII (Readability: 10 g) [Permanent - 600 kg] | 0 to 600 kg | ± 30 g | Using E2, F1, F2, M1 Class Weights by Comparison Method as per OIML R 76-1 |
| Spring Balance (Readability: 10 g) [Permanent] | 0.2 g to 50 kg | ± 2.9 g | Using Differential Newton Weights as per IS 16514-1:2024 |
| Spring Balance (Readability: 100 g) [Permanent] | > 50 kg to 200 kg | ± 28.89 g | Using Differential Newton Weights as per IS 16514-1:2024 |
| Standard Reference Weights - Accuracy Class E1 & Coarser (1 mg to 2 g) | 1 mg, 2 mg, 10 mg, 20 mg, 100 mg, 1 g, 2 g | 1 mg: ± 0.001 mg | 10 mg: ± 0.001 mg | 100 mg: ± 0.0015 mg | 1 g: ± 0.003 mg | 2 g: ± 0.004 mg | Using E1 Class Weight and Electronic Micro Balance (Readability: 1 µg) by Subdivision Method (ABBA Cycle) as per OIML R 111-1 |
| Standard Reference Weights - Accuracy Class E1 & Coarser (5 g to 50 g) | 5 g, 10 g, 20 g, 50 g | 5 g: ± 0.0051 mg | 10 g: ± 0.0061 mg | 20 g: ± 0.007 mg | 50 g: ± 0.01 mg | Using E1 Class Weight and Electronic Micro Balance (Readability: 0.01 mg) by Subdivision Method (ABBA Cycle) as per OIML R 111-1 |
| Standard Reference Weights - Accuracy Class E1 & Coarser (100 g & 200 g) | 100 g, 200 g | 100 g: ± 0.015 mg | 200 g: ± 0.05 mg | Using E1 Class Weight and Electronic Micro Balance (Readability: 0.01 mg) by Substitution Method (ABBA Cycle) as per OIML R 111-1 |
| Standard Reference Weights - Accuracy Class E2 & Coarser (2 kg & 5 kg) | 2 kg, 5 kg | 2 kg: ± 1 mg | 5 kg: ± 1.7 mg | Using E1 Class Weight and Electronic Precision Balance (Readability: 0.001 g) by Substitution Method (ABBA Cycle) as per OIML R 111-1 |
| Standard Reference Weights - Accuracy Class F1 & Coarser (500 g & 1 kg) | 500 g, 1 kg | 500 g: ± 0.82 mg | 1 kg: ± 0.9 mg | Using E1 Class Weight and Electronic Precision Balance (Readability: 0.001 g) by Substitution Method (ABBA Cycle) as per OIML R 111-1 |
| Standard Reference Weights - Accuracy Class F2 & Coarser (20 kg) | 20 kg | ± 100 mg | Using E1 & F1 Class Standard Weight and Electronic Precision Balance (Readability : 0.1 g) by Substitution Method (ABBA Cycle) as per OIML R 111-1 |
Frequently Asked Questions
Are Mass - Weights & Balance calibrations NABL accredited?
Yes. Our Mass - Weights & Balance laboratory is accredited under ISO/IEC 17025:2017 by NABL (Certificate No: CC-4178), ensuring global acceptance and unbroken traceability to NPL/NIST primary standards.
Do you offer on-site Mass - Weights & Balance calibration at client plants?
Yes, Reva Calibration Laboratories deploys mobile metrology teams across Chennai, Tamil Nadu, and PAN-India industrial hubs for on-site Mass - Weights & Balance calibration, minimizing equipment downtime.
What is the standard turnaround time for calibration certificates?
Standard turnaround time is 3 to 5 working days for permanent laboratory orders. Express same-day and next-day calibration slots are available upon prior request.
How do I request a quotation for Mass - Weights & Balance?
Click the "Request a Quote" button on this page to pre-select Mass - Weights & Balance, upload your instrument list (up to 15 MB PDF/Excel), or call +91 98406 72352 for immediate scope verification.