Skip to main content
Traceable Technical Service Scope Permanent Lab Calibration + On-Site Field Calibration

NDT Equipment Calibration Services

Ultrasonic Flaw Detectors, Thickness Gauges, Magnetic Particle Yokes, UV Lamps & Eddy Current Meters.

WhatsApp Lab Desk
Calibration Capability (CMC)
Linearity: ± 1.0% | Velocity: ± 5 m/s | Thickness: ± 0.01 mm.
Primary Master Equipment
IIW V1 & V2 Calibration Blocks, Step Blocks (1 mm to 50 mm), Calibrated Gaussmeters, Calibrated Digital Radiometer Photometers, Artificial Defect Specimens.
Governing Standards
ASME Sec V, EN 12668-1, ASTM E1444, ASTM E1417.
Key Parameters & Applications
Time base linearity, vertical linearity, amplifier gain accuracy, magnetic field lifting power (min 4.5 kg AC / 18 kg DC), UV-A intensity (min 1000 µW/cm²).

Instruments & Parameters Covered (3)

Search or review approved measurement ranges and Calibration & Measurement Capabilities (CMC).

Instrument / Parameter Calibration Range CMC (±) Master / Method
Ultrasonic Flaw Detectors (UT) & Phased Array Frequency 0.5 MHz to 25 MHz (Sound path up to 10,000 mm) Horizontal: ± 0.5% | Vertical: ± 1.0 dB Time base horizontal linearity, receiver vertical linearity, calibrated gain attenuator (dB)
Ultrasonic Digital Thickness Gauges 0.5 mm to 500 mm in steel ± 0.01 mm Sound velocity calibration, zero offset probe delay, measurement linearity over stepped range
Magnetic Particle Inspection (MPI) Electromagnetic Yokes AC & DC Articulated Leg Yokes (Lifting capacity 4.5 kg to 25 kg) Lifting weight verified to ± 0.01 kg AC magnetic lifting force (min 4.5 kg / 10 lb), DC lifting force (min 18 kg / 40 lb), tangential field

Frequently Asked Questions

Are NDT Equipment calibrations NABL accredited?

Our NDT Equipment services follow recognized ASTM, ASME, ISO 9001, and industry guidelines with traceable calibration masters, maintained distinctly from our formal NABL CC-4178 calibration schedule.

Do you offer on-site NDT Equipment calibration at client plants?

Yes, Reva Calibration Laboratories deploys mobile metrology teams across Chennai, Tamil Nadu, and PAN-India industrial hubs for on-site NDT Equipment calibration, minimizing equipment downtime.

What is the standard turnaround time for calibration certificates?

Standard turnaround time is 3 to 5 working days for permanent laboratory orders. Express same-day and next-day calibration slots are available upon prior request.

How do I request a quotation for NDT Equipment?

Click the "Request a Quote" button on this page to pre-select NDT Equipment, upload your instrument list (up to 15 MB PDF/Excel), or call +91 98406 72352 for immediate scope verification.

Explore Related Calibration Disciplines & Services

Industrial Turnkey AMC Plans • NABL Lab Setup Consultancy • ISO/IEC 17025:2017 Training Courses • Direct Lab Helpdesk